| Dimension and Weight | Dimension: 310W x 440D x 1380H(mm) (Excluding IV pole and status indicator) Installation space: 495W x 655D (mm) Weight: Approximately 75kg (Without option) |
|---|---|
| Power Supply | AC110, 220, 240V 50Hz/60Hz 1.3kVA (~2.1kVA) Blood Pump back-up battery |
| Water Supply | Water pressure: 69 to 300 kPa at water flow rate more than 1000 mL/min Water temperature: 5 to 30 ℃ |
| Dialysate | Dialysate concentration: 130 to 159 mmol/L Bicarbonate concentration: 20 to 35 mmol/L Na profile: 130 to 180 mmol/L Flow rate: 300 to 600 mL/min Temperature: 34 to 39.9 ℃ |
| Ultrafiltration | System: Closed loop circuit + UF pump UF rate: 0 to 5.00 L/h UF profile: 0 to 150 % Accuracy: ±30 mL/h |
| Blood Pump | Flow rate: 0, 50 to 600 mL/min (ID8.5, OD12mm tube) 0, 30 to 360 mL/min (ID6.4, OD9.6mm tube) Accuracy: ±20 mL/h |
| Syringe Pump | Syringe size: 10, 20 and 30 mL Flow rate: 0, 0.3 to 9.9 mL/h (0.1 mL step) with bolus function |
| Monitoring System (with alarm setting) |
Venous pressure: -100 to +400 mmHg Arterial pressure: -200 to +400 mmHg TMP: -100 to +500 mmHg Dialysate pressure: -500 to +300 mmHg Blood leak detector: Optical sensor, 500 ppm (Hct 25%) Air detector: Ultrasonic, 0.07 to 0.5 mL at blood flow of 200 mL/min |
| Rinsing/Disinfection | Chemical disinfection and acid rinsing with automatic rinse Disinfectant: 4 to 6 % sodium hypochiorite recommended Dilution ratio 1:10, 1:35 or 1:64 can be selected. Acid rinse: 30% acetic acid recommended Dilution ratio 1:10, 1:20 or 1:30 can be selected. |
| Options | Substitution fluid pump Arterial line air detector Blood pressure manometer Dialysate filter systems Hot rinse system (110V compliant machine is not equipped with hot rinse system) PBC (Powder bicarbonate cartridge ) |