Comparison of performance between NanoAct ™ and colloidal gold
The following two studies were conducted by DCN Diagnostics using identical materials in an optimized full strip format.
NanoAct ™ showed greater LOD (limit of detection) and smaller %CV.


*n=10 in each concentration , *Test Lines (after 15 minutes) were compared by reader (mV), *LOD : Limit of Detection,
*%CV = σ/Ave. ×100 (%),
*materials: antibodies, conjugate pads, nitrocellulose membranes, absorbent pads, and sample pads (Study 1), blood separation filters (Study 2)
NanoAct ™ showed a detection limit of 0.03 ng/mL of Troponin I (TnI).


*n=10 in each concentration , *Test Lines (after 15 minutes) were compared by reader (mV), *LOD : Limit of Detection,
*%CV = σ/Ave. ×100 (%),
*materials: antibodies, conjugate pads, nitrocellulose membranes, absorbent pads, and sample pads (Study 1), blood separation filters (Study 2)
※The results presented on this page are just those from model studies and provide no guarantee of performance.