Hemodialyzer / APS-H: Hemodiafilter - Best suitability for HDF.

Performances

Clearance (in vitro)

APS-18H APS-21H
Blood Flow (mL/min) Qb 200 300 300 400 200 300 300 400
Qf 0 0 75* 100* 0 0 75* 100*
Qd 500 500
Clearances (mL/min)1) Urea 190 251 271 325 192 257 272 326
Creatinine 181 231 264 316 188 249 264 316
Phosphate 178 225 251 301 183 235 255 306
Vitamin B12 145 170 196 233 156 187 205 244
Inulin 99 101 138 161 115 125 161 189
KoA (Urea)(mL/min)2) 836 916
KUF (mL/hr/mmHg (mL//hr/kPa))3) 86 (645) 94 (705)
Effectives Surface Area (m2) 1.8 2.1
Sieving Coefficient4) Inulin 1.0
β2-microglobulin 0.8
Myoglobin 0.4
Albumin < 0.001

Conditions:
1) KoA: Qb=300mL/min, Qd=500mL/min, Qf=0mL/min
2) KUF: bovine blood, TP=60±5g/L, Hct=32±3%, Qb=300mL/min
3) Sieving Coefficient: bovine plasma, TP=60±5g/L
ISO 8637:2010
*post-dilution HDF

Trademark
APS is a trademark of Asahi Kasei Medical Co., Ltd.

Contact Us

Please feel free to send us any questions you may have about our products and support.

Contact Us

Your Voice

Share your feelings and experiences when using our products.

Your Voice
Note: It might be required in your country that medical devices shall be officially registered and/or approved by an authorized organization. Please contact our representative in your region or contact us to confirm whether the listed products are approved for use. These information may contain general information on medical-related topics and is not meant to provide specialist advice or instructions concerning the products and services sold by Asahi Kasei Medical. Please contact us directly for special inquiries. You should not use this information for diagnosing a health problem or disease but should always consult your own physician.Contact us