- Features
- Performances
- Options
- Specifications
- References
![](/medical/en/apheresis/product/equipment/sigma/img/img_12.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_13.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_14.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_15.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_16.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_17.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_18.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_19.jpg)
![](/medical/en/apheresis/product/equipment/sigma/img/img_20.jpg)
Treatment Mode | CRRT | CVVH, CVVHD, CVVHDF, SCUF | ||
---|---|---|---|---|
Apheresis | PE, PA, DFPP, LCAP, HA | |||
Blood Pump | Type | 3-roller tubing pump | ||
Tube setting | Movable stator (interlocked with cover) | |||
Flow rate control range | 1 to 400 mL/min (CRRT) | |||
Flow rate control range | 1 to 200 mL/min (Apheresis) | |||
Pump (FP) | Type | 3-roller tubing pump | ||
Tube setting | Movable stator (interlocked with cover) | |||
Flow rate control range | 0.01 to 12.00 L/h | |||
Pump (DP) | Type | 3-roller tubing pump | ||
Tube setting | Movable stator (interlocked with cover) | |||
Flow rate control range | 0.01 to 10.00 L/h | |||
Pump (RP) | Type | 3-roller tubing pump | ||
Tube setting | Movable stator (interlocked with cover) | |||
Flow rate control range | 0.01 to 10.00 L/h | |||
Syringe Pump | Applicable syringe | 20, 30 and 50 mL syringes | ||
Flow rate control range | 0.1 to 15.0 mL/h | |||
Fast feed flow rate | 0.1 mL/s or more | |||
Air Detector 1 | Type | Ultrasonic transmission | ||
Sensitivity | 0.1 mL or more (blood pump flow rate: 100mL/min) | |||
Air Detector 2 | Type | Ultrasonic transmission | ||
Sensitivity | 0.01 mL or more (blood pump flow rate: 100mL/min) | |||
Blood Detector | Type | Optical (infrared) | ||
Empty Detector 1 | Type | Ultrasonic transmission | ||
Empty Detector 2 | Type | Ultrasonic transmission | ||
Blood Leak Detector | Type | Optical (Infrared) | ||
Pressure Sensors |
Arterial Pressure | Measurement range | - 500 to 500 mmHg | |
Inlet Pressure | Measurement range | - 500 to 500 mmHg | ||
Venous Pressure | Measurement range | - 500 to 500 mmHg | ||
Filtrate Pressure | Measurement range | - 500 to 500 mmHg | ||
2nd Filter Pressure | Measurement range | - 500 to 500 mmHg | ||
Heater 1 | Type | Single-sided heating plate | ||
Temperature set range | 35 to 40℃ (Plate surface) | |||
Heater 2 | Type | Single-sided heating plate | ||
Temperature set range | 35 to 40℃ (Plate surface) | |||
Scale | Rating | 3 kg | ||
Battery | Operating time | Can operate blood pump for about 15 min | ||
Electrical Ratings | Rated voltage | AC 100 to 240 V | ||
Rated frequency | 50/60 Hz | |||
Power unit | 400 VA | |||
Signal I/O P1 | Remote alarm port | |||
Signal I/O P2 | Ethernet port | |||
Electric Shock Protection | Class I equipment, Type BF applied part | |||
Waterproofing | IPX1 | |||
EMC | Conform to EMC standard EN60601 1 2: 2001 | |||
Dimensions | 496[W] X 1454[H] X 547[D] mm | |||
Weight | Approx.70 kg (including options) |
Trademark
Plasauto is a trademark of Asahi Kasei Medical Co., Ltd.
![](/medical/en/apheresis/product/equipment/sigma/img/img_download.jpg)
PlasautoΣ
Please feel free to send us any questions you may have about our products and support.
Share your feelings and experiences when using our products.